Troubleshooting
Efficia CM Series Service Guide 4-19
Managing Component Usage Parameters
Managing Component Usage Parameters
The monitor tracks the usage of some components and displays them on the
Diagnostics Menu. This information can help to diagnose monitor problems.
The monitor tracks the following parameters:
Note — CO
2
usage information is displayed in the CO2 Test menu, accessed through
the Diagnostics Menu. See Chapter 3, “Performance Verification Testing.”
Resetting Component Usage Parameters
Caution Always reset a component usage parameter after replacing the associated part.
To reset the parameter:
Parameter Description
NBP Cycle Count
The number of NBP measurements taken by the monitor.
LCD Usage Hours
The number of hours that the LCD screen has been on.
Step
1
Open the Maintenance options of the Diagnostics Menu. See “Accessing
the Diagnostics Menu and Maintenance Options” on page 3-6.
When Maintenance options are available, a Reset button appears next to
each parameter.
2
Select the Reset button to the right of the parameter to reset.
A confirmation window appears before the parameter is cleared.
3
In the confirmation window, select OK to reset the parameter to 0.
4
Repeat step 2 and step 3 to reset another parameter.
5
Close the menu.