Performance Verification Testing
Efficia CM Series Service Guide 3-3
Required Test Equipment
“Oridion Sidestream CO
2
Calibration Test” on page 3-11
Electronic flowmeter, M1026-60144
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
“Respironics CapnoTrak Sidestream
CO
2
Accuracy Test” on page 3-14
Respironics calibration gas regulator,
Respironics PN 6081-00.
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
“Respironics LoFlo Sidestream CO
2
Accuracy Test” on page 3-16
Respironics CO
2
airway adapter.
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
“Mainstream CO
2
Accuracy Test”
on page 3-18
• CO
2
sensor
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
M1990A airway adapters or similar (2)
“ECG/Respiration Performance
Test” on page 3-19
Patient simulator
ECG patient cable
“ECG Output Performance Test” on
page 3-23
Defibrillator with Defib Sync cable
Note — Your defibrillator must be capable of
accepting digital output from the monitor.
Patient simulator
“NBP Test” on page 3-20 A reference manometer (including hand
pump and valve), with an accuracy of 0.2%
An expansion chamber (volume 250 ml ±
10%)
Tubing
“IBP Test” on page 3-22 Patient simulator
“Temperature Test” on page 3-24 Patient simulator (with an accuracy of 0.1
o
C
or 0.2
o
F)
To Perform This Test You Need This Test Equipment