Managing Logs
Troubleshooting
4-22 Efficia CM Series Service Guide
To delete the logs:Clearing Patient Data
Patient data and a summary are automatically stored in internal memory. A summary
report can be downloaded to a USB flash drive.
If you return a monitor for service, you must first clear the patient data. The Clear Data
function resets the user settings to the factory defaults and clears the following:
• All patient data
• All profiles and profile settings
• All logs
• All cloneable settings, except the AC line frequency
• Monitor name
The following settings are not cleared:
• Monitor serial number
• Key code
• LAN/WLAN MAC Address
• Touchscreen, temperature, IBP, and C.O. calibration data
• Language
• NBP cycle count
• LCD usage time
• Time zone
• AC line frequency
• RTC calibration data
• Country code
3
Select the View Logs button.
The View Log Menu appears.
4
Select the Manage Logs tab.
5
Select the Save All button.
All logs are saved to the USB flash drive as a tar file named: SN_logs.tar,
where SN is the serial number of the monitor.
6
Close the menu.
1
Access the Admin tab. See “Accessing the Admin Functions” on page 3-5.
2
Select the View Logs button.
The
View Log Menu appears.
3
Select the Manage Logs tab.
4
Select the Clear All button.
All logs are deleted from the monitor.
5
Close the menu.