Efficia CM Series Rack and Modules Instructions for Use 9-3
Electrical and Environmental Specifications
Battery Run Time
(Monitoring ECG, SpO
2
and
15 minute NBP interval
measurements, 25°C, LCD
screen brightness at 6)
With the monitor connected to the rack and with one fully
charged 9 cell battery.
> 1 hour for CM12, CM120, and CM150 monitors.
With the monitor connected to the rack and with two fully
charged 9 cell batteries.
> 3 hours for CM12, CM120 and CM150 monitors.
See “Rack Power” on page 9-6 for more information.
Mechanical Shock The monitor, rack and modules operate after exposure to
the following mechanical shock standards.
Operating:
Half-sine waveform, Peak Acceleration 50g’s.
Velocity of < 55 in/sec
Pulse duration < 3 ms, one impact on each of six
faces.
Non-Operating Survival:
Trapezoidal waveform, acceleration >= 30g
Faired Acceleration velocity change = 742 cm/s +/-
5%
One impact on each of six faces.
Test conditions include:
Peak Acceleration: 150 m/s2 (15.3g)
Duration: 11 ms
Pulse shape: half sine
Number of shocks: 3 shocks per direction per axis (18
total)
Mechanical Vibration Test conditions include:
Frequency range: 10 Hz to 2000 Hz
Resolution: 10 Hz
Acceleration amplitude:
10 Hz to 100 Hz 1.0 (m/s2)2/Hz.
100 Hz to 200 Hz -3 dB/octave
200 Hz to 2000 Hz 0.5 (m/s2)2/Hz
Duration: 10 min per each perpendicular axis (3 total)
Rack to monitor Interface
Supply Voltage range
9-15V
The maximum current draw through rack is 3.6A while
Gas, dual IBP and BIS modules are inserted.
Thermal
Operating
Temperature
1
Monitor, rack, and modules:
•10
o
C to 40
o
C (50
o
F to 104
o
F)
Parameter Specification