Electrical and Environmental Specifications
Specifications
9-4 Efficia CM Series Rack and Modules Instructions for Use
Storage
Temperature
2
Monitor, rack, and modules:
• -20
o
C to 50
o
C (-4
o
F to 122
o
F)
Monitor with recorder paper
• -20
o
C to 40
o
C (-4
o
F to 104
o
F)
Dräger OEM Gas module:
• -20
o
C to 75
o
C (-4
o
F to 167
o
F)
Covidien BISx4 patient cable:
• -40
o
C to 70
o
C (-40
o
F to 158
o
F)
Oridion Microstream™ CO
2
:
• -40
o
C to 70
o
C (-40
o
F to 158
o
F)
Transport
Temperature
-20
o
C to 40
o
C (-4
o
F to 104
o
F)
Humidity
Operating Humidity
3
Monitor, rack, and modules:
• 15% to 90% RH, without paper non-condensing
Storage Humidity 15% to 90% RH, non-condensing
Transport Humidity 15% to 90% RH, non-condensing
Atmospheric Pressure
Operating
Atmospheric
Pressure
4
Monitor, rack, and modules:
• 0 m to 3,000 m (0 ft to 9,842 ft) above sea level
• 1,013 mbar to 701 mbar (101 kPa to 70 kPa)
Storage Atmospheric
Pressure
Monitor, rack, and modules:
• 0 m to 3,000 m (0 ft to 9,842 ft) above sea level
• 1,013 mbar to 701 mbar (101 kPa to 70 kPa)
Transport
Atmospheric Pressure
701 hPA to 1013 hPA
Electromagnetic
Compatibility
The monitor, rack and modules meet the IEC 60601-1-2:2014,
4th Edition Standard.
Meets IEC Class A emission levels
1. Composite test range: 10° C to 40° C
2. Composite test range: -20° C to 50° C without paper
3. Composite test range: 15% to 90% without paper non-condensing
4. Composite tested 1013 - 701 mbar, 0 - 3000 meters, 0 - 9842 feet above sea level.
Parameter Specification