ErgoLine D330/D340
Customer Engineer Manual
Page 4-4
Note
: Codes 57 and 58 are used in conjunction with the other codes.
It is essential that forced transmit or receive is deactivated off before
returning to user mode. See Note
1
on page 4-4.
Digital Interfaces (Codes 61… 70)
Code 61 : tests Y signal 0 on PC interface
Code 62 : tests Y signal 1 on PC interface
Code 63 : sends a signal from the PC interface at 1200 baud
Code 64 : tests X interface (activates X interface mirrored mode)
Code 65
1
: tests static interface signal 0
Code 66
1
: tests static interface signal 1
Code 67
1
: tests V24
Code 68
1
: tests line interface info_0
Code 69
1
: tests line interface info_1
Code 70
1
: tests B channel loopback (only when 2 B-channels available)
Note
1
: It is recommended that after any of these tests, a hardware reset is
completed.