Exposure parameters for conditioning
Tube adapted # Exposures Tube not adapted
kV mA ms kV mAs
80 10 50 < 1 > 80 0.5
80 10 500 < 1 > 80 5
80 200 250 < 1 > 80 50
10 seconds pause 10 seconds pause
80 max. mA 100 < 1 > 80 100
1 minute pause 1 minute pause
90 10 50 < 1 > 90 0.5
90 10 500 < 1 > 90 5
90 200 250 < 1 > 90 50
10 seconds pause 10 seconds pause
90 max. mA 100 < 1 > 90 100
1 minute pause 1 minute pause
100 10 50 < 1 > 100 0.5
100 10 500 < 1 > 100 5
100 200 250 < 1 > 100 50
10 seconds pause 10 seconds pause
100 max. mA 100 < 1 > 100 100
1 minute pause 1 minute pause
110 10 50 < 1 > 110 0.5
110 10 500 < 1 > 110 5
110 200 250 < 1 > 110 50
10 seconds pause 10 seconds pause
110 max. mA 100 < 1 > 110 100
1 minute pause 1 minute pause
120 10 50 < 1 > 120 0.5
120 10 500 < 1 > 120 5
120 200 250 < 1 > 120 50
10 seconds pause 10 seconds pause
120 max. mA 100 < 1 > 120 100
1 minute pause 1 minute pause
130 10 50 < 1 > 130 0.5
130 10 500 < 1 > 130 5
130 200 250 < 1 > 130 50
10 seconds pause 10 seconds pause
130 max. mA 100 < 1 > 130 100
1 minute pause 1 minute pause
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
48 CSIP Level 1 (08.0)
© 2008 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Conv test Optimus