Exposure parameters for conditioning
Tube adapted # Exposures Tube not adapted
kV mA ms kV mAs
140 10 50 < 1 > 140 0.5
140 10 500 < 1 > 140 5
140 200 250 < 1 > 140 50
10 seconds pause 10 seconds pause
140 max. mA 100 < 1 > 140 100
1 minute pause 1 minute pause
145 10 50 < 1 > 145 0.5
145 10 500 < 1 > 145 5
145 200 250 < 1 > 145 50
10 seconds pause 10 seconds pause
145 max. mA 100 < 1 > 145 100
1 minute pause 1 minute pause
148 10 50 < 1 > 148 0.5
148 10 500 < 1 > 148 5
148 200 250 < 1 > 148 50
10 seconds pause 10 seconds pause
148 max. mA 100 < 1 > 148 100
1 minute pause 1 minute pause
150 10 50 < 1 > 150 0.5
150 10 500 < 1 > 150 5
150 200 250 < 1 > 150 50
10 seconds pause 10 seconds pause
150 max. mA 100 < 1 > 150 100
1 minute pause 1 minute pause
Converter test kit OPTIMUS for OPTIMUS 50/65/80 gen‐
erators release 3.x with converters 4512 104 7231x
Conv test Optimus CSIP Level 1 (08.0)