EasyManua.ls Logo

Philips Pathology Scanner SG20 - Page 28

Philips Pathology Scanner SG20
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Preparation of slides
28
4522 207 12671/ * 2021-06-17
Pathology Scanner SG20 / SG60 / SG300
• undamaged
• of uniform thickness
• not torn
Fig. 11: Cutting the tissue
Placing tissue on the slide
• To minimize scan time, place tissue vertically on the glass slide.
For slides containing multiple tissue sections: if there is 0,5-1 mm distance between pieces
of tissue, the multiple small areas will be scanned.
Fig. 12: Alignment of tissue
• Keep 3 mm between the tissue and the slide edges and/or markers.
• Keep 5 mm between the tissue and the barcode area or slide brand text.

Table of Contents