Operation Preparation of slides
28
4522 207 12671/ * 2021-06-17
Pathology Scanner SG20 / SG60 / SG300
• undamaged
• of uniform thickness
• not torn
Fig. 11: Cutting the tissue
Placing tissue on the slide
• To minimize scan time, place tissue vertically on the glass slide.
For slides containing multiple tissue sections: if there is 0,5-1 mm distance between pieces
of tissue, the multiple small areas will be scanned.
Fig. 12: Alignment of tissue
• Keep 3 mm between the tissue and the slide edges and/or markers.
• Keep 5 mm between the tissue and the barcode area or slide brand text.