Planned maintenance Maintenance
59
4522 207 12671 * 2021-06-17
Pathology Scanner SG20 / SG60 / SG300
Item What to check Frequency
Earth (ground) Check earth (ground) of scanner Yearly or after 540,000 slides
Mechanical Check slide transport Yearly or after 540,000 slides
Controls and indicators Check accuracy and function of
the following:
- all controls
- all visible/audible indicators
Yearly or after 540,000 slides
Image quality Check performance Yearly or after 540,000 slides
Labels Check for legibility Yearly or after 540,000 slides
Tab. 15: Planned Maintenance Frequencies