Troubleshooting Power failure
72
4522 207 12671/ * 2021-06-17
Pathology Scanner SG20 / SG60 / SG300
NOTICE
Using this pin during normal operation of the scanner will result in an error.
When the power returns:
Use the power button on the front side of the scanner to switch the scanner on.
Start (or restart) the scanning process by closing the door.
NOTICE
After a power failure, a slide can still be present on the stage. This can cause a workflow error
during start-up of the scanner.
To make sure the scan-status of this slide is correct and complete, remove the slide from the
stage, place it in a rack and scan it again. See chapter “Remove slide from stage area” on page
86.