Chapter 8
Alarms and messages
1047358 Rev N Respironics V60 Ventilator User Manual 8-11
Table 8-3: Check Vent alarm messages: summary and troubleshooting
Message Description Corrective Action
Priority
type (ID)
Manually
resettable
Autore-
settable
Silence-
able
Check Vent: 1.8 V
Supply Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(21)
Yes No No
Check Vent: 3.3 V
Supply Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(22)
Yes No No
Check Vent: 5 V Sup-
ply Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(23)
Yes No No
Check Vent: 12 V
Supply Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(24)
Yes No No
Check Vent: 24 V
Supply Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(24)
Yes No No
Check Vent: 35 V
Supply Failed
Technical failure Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(26)
Yes No No
Check Vent: Air Flow
Sensor Calibration
Data Error
Flow-related patient data is
disabled. Oxygen concentra-
tion switches to 21% (venti-
lates with air only). Default
volume used in AVAPS
mode. Standby disabled.
Volume, leak, disconnect,
and occlusion alarms com-
promised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(13)
Yes No No
Check Vent: Alarm
LED Failed
Technical failure. Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High (6) Yes No No
Check Vent: Auxilia-
ry Alarm Supply
Failed
Backup alarm problem Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(20)
Yes No No
Check Vent: Backup
Alarm Failed
Backup alarm problem Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High (5) Yes No No
Check Vent: Barome-
ter Calibration Data
Error
Default barometric pres-
sure of 686.0 mmHg (ap-
proximately 900 m/2953 ft
above sea level) used in cal-
culations
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(18)
Yes No No