74
7.3.4 GAS 150
Warm-up time: 45 minutes
Input/Output Connector: Parker Prestolock 6 mm
Weight: 5 kg
Dimensions: 41 x 37 x 17 cm
Carbon Dioxide Module Specification:
Available CO
2
ranges: 0 2 %, 0 3 %, 0 5 %, 0 10 %, and 0 20 % (please note that if any other than standard range is
required, the customer needs to specify the required range)
Standard Range: 0 2 %
Accuracy at +25 °C (+77 °F) against certified factory references: <±[0.02% CO
2
+ 2% of reading]
Nonlinearity: <±1.0 %FS
Temperature dependence of output: (typ.) 0.1 %FS/°C (0.1 %FS/°F)
Pressure dependence: typically 0.15% of reading/hPa
Long-term stability: <±5 %FS/2 years
Response time (63 %): 20 seconds
Flow meter/controller Specification:
Range: based on customer´s specifications
Default range: 1 l/min (0.02 1 l/min)
Accuracy (incl. linearity): ± 1% FS plus ± 1% RD
Repeatability: < 0.2% FS typical
Settling time: approx. 2 seconds
Temperature sensitivity: zero: < 0.1% FS/°C; span: < 0.2% RD/°C
Pressure sensitivity: 0.3% RD/bar typical at Air
Attitude sensitivity: max. error at 90° off horizontal 0.2 % at 1 bar
Input Pressure: 3 bar