© PSI (Photon Systems Instruments), spol. s r. o.
28
(4) Inlets for the gas dispenser tubing.
(5) Connectors for optional sensors closed with filler screws.
(6) Sampling tube for turbidostat mode. Inlet is used to add medium to the vessel or to collect the samples.
(7) Excess medium outlet.
(9) Inlets for chemostat mode. Inlets for base and acid chemostat mode.
(10) Optional inlets (e.g. for multiple peristaltic pump connection).
Fig. 15 Components of 3000 ml cultivation vessel.
4.3 PH/TEMPERATURE SENSOR
The pH/temperature module is designed for continuous measuring of pH and temperature in the medium. pH/temperature
sensor is part of minimal required configuration (described in chapter 3.2) and is recommended for purchase if precise
control of cultivation conditions is required.
The module consists of electronic unit embedded in Photobioreactor FMT150 Control Unit, probe and communication
cable (Fig. 21).
IMPORTANT NOTE: It is strongly recommended to avoid autoclaving pH electrode. The manufacturer
states that autoclaving deteriorates it rapidly. Use warm water with a detergent or chemical
disinfectant for its cleaning.