When the reminding prompt appears, the red light of the reminding indicator will be
on. When the reminding is released, the red light turns off.
8.3 Reminding function
The Medical Diode Laser monitors the operating status of the system in real time.
through Sounds, indicators, and graphical warning signs alert user happened
unusual events and interrupting the laser laser outputting.
1.Laser temperature exceeds
45 ° C;
2. Fiber coupler temperature
exceeds 60
℃
;
3. Temperature sensor failure;
1. Turn off the power, if the temperature
reminding caused by the high temperature,
air cooling device for 10 minutes and it will
return to normal after the boot;
2. Please contact the company after-sales
maintenance.
Laser temperature is below 0℃;
1. Turn off the power;
2.Raise the ambient temperature;
The remote interlock is not
inserted correctly or is not
plugged into the device;
1.Check the rear panel of the device and
whether the remote interlock is correctly
inserted into the device;
2. If the remote interlock is inserted into the
device and the reminding condition is still
unresolved, please contact the company
for after-sales maintenance.
1.The foot switch is not properly
inserted or is not inserted into
the device;
2.Press the handle switch button
when it’s not into the "Ready"
state;
1.Check the front panel and whether the
handpiece switch is correctly inserted into
the device;
2.If the handle switch is properly inserted
into the device, the reminding status is still
unresolved, please check the handle
switch button is pressed or not;
3. Check the above operation, if there is
still reminding status, please contact the
company after-sales maintenance.
1.Fiber connector is not inserted
into the laser window;
2.Micro - switch of laser window
fails;
1.Insert the fiber connector into the laser
window of the device;
2.If there is still an reminding condition,
please contact the company after-sales
maintenance.
The emergency stop button was
pressed.
Press the emergency stop button again.