Planmeca Proline XC Panoramic X-ray 15
GRAPHIC USER INTERFACE
User's manual
Adjustment range of AEC exposure parameter
Child: the range varies from -4 to +3 units compared
to the original values.
Small or average patient: the range varies from -5 to
+5 units compared to the original values.
Large patient: the range varies from -5 to +6 units
compared to the original values.
One unit corresponds to 1 mA or 2 kV, i.e. the
maximum increase of the exposure parameter for
example for child exposure can be 3 mA and 6 kV or
2 mA and 2 kV.