Planmeca ProX 15
CONTROL PANEL
User’s Manual
8.2.3 SELECT key
Press the SELECT key briefly to select the parameter - kV,
mA or exposure time - to be changed. When the parameter
value is flashing on the display, the parameter can be
changed. After adjusting the kV or mA value or exposure
time the unit will return automatically to the time
adjustment mode after 5 seconds time.
Press and hold down the SELECT key (about 4 seconds)
until you have heard two signal tones to enter the
programming mode. For more information about
programming refer to chapter 16.1 “Programming default
exposure values” on page 36.
Press the SELECT key to clear the error from the display.
8.2.4 MODE key
To select the exposure parameters for film, digital or
phosphor plate imaging press and hold down the MODE
key for 2 seconds.
To enter the digital imaging mode from the film-based
imaging mode press and hold down the MODE key for 2
seconds. There is no prefix on the display in the film-based
imaging mode. The exposure time with prefix d. appears
on the time display in the digital imaging mode. All the keys
function as in the film-based imaging mode.
To enter the phosphoric mode from the digital imaging
mode press the MODE key and hold it down for 2 seconds.
The exposure time with prefix P. appears on the time
display in the phosphoric mode. All the keys function as in
the film-based imaging mode.
All the exposure parameters remain selected after the
exposure until the user changes the parameters or until the
unit is switched off. The selected mode stays in the unit
memory even in case of power failure.
If the DAP display is activated (service mode parameter
24=2 or 3) by pressing the MODE key briefly the DAP
value mGy*cm2 appears on the time display, the beam
limiting device value between 1 and 12 appears on the kV
display and the letter A on the mA display. The beam
limiting device value can be changed using the arrow up/
down keys.
If the exposure count display is activated (service mode
parameter 23=2) by pressing briefly the MODE key the
exposure count value (00000 - 50000) appears on the mA
and time displays. The text EC appears on the kV display.
To return to the exposure value (kV, mA, sec) displays
press briefly the SELECT key.
Film-based
imaging mode
Digital
imaging mode
Phosphoric
imaging mode

Table of Contents

Other manuals for Planmeca ProX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProX and is the answer not in the manual?

Planmeca ProX Specifications

General IconGeneral
TypeIntraoral X-ray
Tube current7 mA
Imaging TechnologyDigital
Focal spot0.4 mm
X-ray generatorHigh-frequency
Power Requirements100–240 V AC, 50/60 Hz
Tube voltage60-70 kV