Rev. 02 (May 2023)
Plasmapp Co., Ltd.
STERLINK mini User Manual
19
for one sterilization cycle. When the sterilization cycle is completed or stopped, the used cassettes
should be handled with wearing protective gloves and must be disposed according to the right
regulation. Please refer to the Chapter 5 featuring “Operation” to have more information on how to
use the cassette.
CAUTION! CHECK EXPIRY DATE.
When you use the sterilant cassettes (STERLOAD™ mini), check the expiry date. If the date
is expired, dispose and use new available sterilant cassettes.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle cancellation
or error occurrence. Hydrogen peroxide liquid may be present on the load or in the chamber.