STERLINK mini User Manual
WARNING! HYDROGEN PEROXIDE IS CORROSIVE.
Concentrated hydrogen peroxide is corrosive to skin, eyes, nose, throat, lungs, and the
gastrointestinal tract. Always wear latex, PVC (vinyl), or nitrile gloves while removing items
from the sterilizer following a cancelled cycle or error occurrence. Following a hydrogen
peroxide may be present.
WARNING! DISPOSAL OF HYDROGEN PEROXIDE.
Hydrogen peroxide is designated as a dangerous and hazardous medical waste by the
American Environment Protection Association and should be disposed of in accordance with
applicable local regulations. (U.S. Waste Disposal Regulations: U.S. EPA 40 CFR 262)
CAUTION! USE MANUFACTURER SUPPLIED POWER CORD ONLY.
To avoid risk of equipment malfunction or electrical shock, use the power cord supp
lied by the manufacturer with the STERLINK mini sterilizer only.
WARNING! DO NOT MODIFY THIS EQUIPMENT.
No modification of this equipment is allowed.
5.2. Start-up and Preheating
1. Turn on the main power switch on the back of the sterilizer.
2. When the power is turned on, the sterilizer will be automatically preheated.
3. Close the door during the preheating process. Preheating may take up to 30 minutes.
4. After the preheating process, the ‘ ’ icon will be displayed instead of ‘Preheating’ as shown below.