EasyManua.ls Logo

Plasmapp STERLINK mini - Cassette Loading

Plasmapp STERLINK mini
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rev. 02 (May 2023)
Plasmapp Co., Ltd.
STERLINK mini User Manual
28
CAUTION! DO NOT REUSE.
The STERLOAD™ mini is intended for single use only. Do not reuse, if you reuse the product
an error will occur.
CAUTION! CHECK THE EXPIRATION DATE.
When you use the STERLOAD™ mini sterilant cassettes, check the expiration date first. If
the date has expired, discard it, and use a new Sterilant Cassette. A cassette that has expired
may cause errors.
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle can
cellation or error occurrence. Hydrogen peroxide liquid may be present on the load o
r in the chamber.
5.4. Cassette loading
The STERLINK mini sterilizer has a mode which is operated by using the cassette of the
STERLOAD™ mini. Place the cassette in the loading block as depicted in the following figure.
Fig. 5.2 Mounting cassette of the STERLOAD™ mini
WARNING! BEWARE OF HOT SURFACES IMMEDIATELY AFTER
STERILIZATION.
Immediately after the sterilization cycle, the interior of the sterilizer may be hot. Never touch
the chamber and the interior of the sterilizer with your bare hands or gloved hands.
WARNING! WATCH OUT FOR HAND INJURIES.
The door of the sterilizer is designed to be opened and closed manually, which may cause
injury to your hands. Be careful not to hurt your hands when opening and closing the door.
This side up

Table of Contents

Related product manuals