Rev. 02 (May 2023)
Plasmapp Co., Ltd.
STERLINK mini User Manual
49
Notes.
A Pre-heating error occurs if the unit is not warmed up to 40 minutes after booting.
An Initializing error occurs if the cassette is re-used, expired, or invalid. Please check cleanliness
of the scanner glass inside the sterilizer chamber and try again with a new cassette.
Initializing error occurs if the chamber door is open.
If the load prepared with too much moisture, an error message can be displayed to stop the
sterilization cycle. The load should be dried and try again with a new cassette.
Many system errors are related to load issues. However, problems can arise due to malfunction of the
STERLINK mini sterilization system. If the problems persist, please contact Plasmapp or your local
Plasmapp customer support representative.
WARNING! POSSIBLE NON-STERILE DEVICE.
Any error may result in a non-sterile medical device, and it requires to restart a new
sterilization cycle with rewrapping using new packaging material.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Any error may result in failure of removing hydrogen peroxide. Wear latex, PVC (vinyl),
or nitrile gloves when handling the loads and cassette.
9.2. Self-Diagnosis Function
Problems can be caused by a malfunction of the STERLINK mini Sterilization System, including user
errors, sterilization cassettes-related errors, and other system errors described above. If you click the
“History” button on the initial screen, you can check the details of the malfunctions (Error Codes) on
the “History” screen and take necessary actions by referring to the Error List below. If the problem
persists, please contact Plasmapp or your local Plasmapp customer support representative.
[ Error List]
CODE
Description
10
Temperature of the chamber and vaporizer does not reach the appropriate range.
11
During the process, an abnormal power outage occurs.