CAUTION! BE CAREFUL! MAY BE CHAMBER HOT!
The temperature in the chamber may be very high. Therefore, when working in
the chamber, be sure to wear protective gloves, and be careful of burns.
Chapter 5. Operation
5.1. Before Use
When using the STERLINK plus Sterilization System, follow the instructions specified in ‘Chapter 4
Preparation of Medical Devices’, and be sure to read and understand the operating instructions and
precautions provided in this user manual.
Note:
▪ This sterilizer should be placed on a table in the designated area for use. It should be installed on a
flat floor (inclination of 3 degrees or less) with a space of at least 5 cm from the back and top
surfaces.
▪ Do not apply unspecified voltage to the sterilizer. If you need to use unacceptable power for urgent
sterilization under special circumstances, please contact your nearest distributor or our head office.
The damages to the product resulting from the use of unspecified voltage will void the warranty.
▪ When preheating the sterilization chamber, be sure to connect the power cord and turn on the power
switch 15 minutes before starting the sterilization process. Starting the sterilization process
without preheating may cause failure.
▪ The LAN connector is used for S/W updates or diagnosing the product status.
▪ Do not use the sterilizer in close proximity to the equipment that generates plenty of
electromagnetic waves.
▪ Keep the sterilizer away from fire or a location where steam is generate
▪ When restarting a sterilizer that has not been used for a long time, a user should contact the
manufacturer to get appropriate instructions.
▪ This equipment can only be operated by people who have been trained by the manufacturer or a
manufacturer-delegated company.