exposed to sterile residues that remain on the cassette. Accordingly, make sure to wear protective
gloves when handling used cassettes.
WARNING! POSSIBLE NON-STERILE DEVICE.
Improper loading of the cassette may result in either a non-sterile device or cycle cancellation.
The cassette should be precisely loaded on the cassette loading block of the sterilizer and
properly locked.
WARNING! ELECTROMAGNETIC DISTURBANCES.
A high voltage is used at the start of the sterilization cycle, which may cause minor
disturbances to electronic devices outside the product. Also, portable RF communications
equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the [ME EQUIPMENT or ME SYSTEM],
including cables specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result in sterilization failure.
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR
11 class B is normally required) this equipment might not offer adequate protection to radio-
frequency communication services. The user might need to take mitigation measures, such as
relocating or re-orienting the equipment.
CAUTION! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle cancellation
or error occurrence. Hydrogen peroxide liquid may be present on the load or in the chamber.