Warning! PRODUCT MANAGER ACCESS AUTHORIZATION
Additional utility functions including the “E-OPEN” functions, which are used when the door
is not opened, is available only to trained administrators of this sterilizer.
Note:
If the drain filter is dirty or blocked by other objects, the draining process will fail, and the door
cannot be opened. When it occurs, this “E-OPEN” can be used to open the door. Note that this
key is activating addition venting and it can take about 10 seconds. If the problems persist, please
contact Plasmapp or your local Plasmapp customer support representative.