7-15
7 Technical Specifications
7.5 Scanning Heads
7.5.1 General Data PSV-I-5xx
Ambient Conditions
ATTENTION!
Damage of the scanning head caused by too high ambient temperature!
If the ambient temperature exceeds 35 °C (95 °F), the scanning head may
overheat. This could damage it.
» At ambient temperatures above 35 °C (95 °F) you must assemble the
scanning head on a tripod (e.g. VIB-A-T02) so that also its bottom side of
the scanning head is ventilated sufficiently.
Operating temperature: +5 °C…+35 °C (41 °F…95 °F) without limitations
+3C+4C (9F10F) with limitations (see
above)
Storage temperature: -10 °C…+65 °C (14 °F…149 °F)
Operating altitude (only
Standard):
max. 3048 m (max. 10000 ft)
Humidity: max. 80%, non-condensing