Safety information
18 / 116
DefiMonitor XD 21324 / GB / K05
by a testing authority licensed to test the device when equipped ready
for use.
Use new, undamaged and unexpired defibrillation electrodes for
every patient to avoid any possible burns to the skin!
Only connect the self-adhesive electrodes with the PRIMEDIC™
DefiMonitor XD. Use of the electrode system with other devices can
lead to dangerous impedance currents for the patient!
Do not use the device in the immediate vicinity of other sensitive
equipment (e.g. measuring equipment that is sensitive to magnetic
fields) or strong sources of interference which could affect the way
the PRIMEDIC™ DefiMonitor XD works. Keep a sufficient distance
away from other therapeutic and diagnostic energy sources (e.g.
diathermy, high frequency surgery, magnetic resonance tomography).
These devices can affect the PRIMEDIC™ DefiMonitor XD and
disrupt the way it operates. For this reason, disconnect all other
devices from the patient before carrying out defibrillation.
Prior to defibrillation, disconnect all other electrically-operated
medical devices that are not defibrillation-proof and are used on the
patient.
Keep the defibrillation electrodes and the ECG electrodes away from
other electrodes, metal objects and earthed objects!
Do not use the device in AED mode on children under the age of 8 or
children with an estimated body weight of less than 25 kg!
Position the electrodes precisely according to the description.
Dry the chest and carefully remove any large amount of hair on the
patient before applying the defibrillation electrodes.
Do not place the defibrillation electrodes directly over an implanted
pacemaker to avoid a possible misinterpretation by the device and to
avoid any damage to the pacemaker from the defibrillation impulse.
Do not touch the patient during the ECG analysis and avoid any
vibrations.
If the ECG analysis is being carried out in a vehicle, the vehicle has
to stop and switch off the engine to guarantee a correct analysis.
Stop any reanimation while the PRIMEDIC™ DefiMonitor XD is
analysing the ECG.
In case of extreme fluctuations in the power supply, the device should
be disconnected from the mains by pulling the mains plug out from its
socket, and the device should be operated using the integrated
PRIMEDIC™ AkuPak battery.
Strong sunlight may make it difficult to read the display.
The use of several medical devices simultaneously may present a