Auto mode (AED mode)
66 / 116
DefiMonitor XD 21324 / GB / K05
In basic state, a dashed line appears on the display and the acoustic message < Check
electrodes > is issued if the electrodes are not connected up. As soon as a reading is activated
via the electrodes, the ECG signal appears on the monitor.
9.4 Defibrillation not required
If the device cannot find a shockable rhythm, then it recommends two minutes of cardio
pulmonary resuscitation (CPR).
< No shock advised >
< Cardiopulmonary resuscitation >
< Give 30 chest compressions >
< Give 2 rescue breaths >
etc.
When the CPR time (defined by the set number of CPR cycles) has elapsed, the device
resumes ECG analysis.
9.5 Keeping the defibrillator ready for use
After a reanimation, clean the device, replace the SavePads, check and if necessary replace
the power supply unit to ensure that the device is ready for operation again as quickly as
possible. Charge the PRIMEDIC™ AkuPak to ensure that sufficient energy is available for the
device to be used again.
If any malfunctions or noticeable problems occur, please contact your nearest service station.