DefiMonitor XD 21324 / GB / K05
6 Device self-test
6.1 Self-test when the PRIMEDIC™ DefiMonitor XD is switched on
The self-test is triggered by switching on the PRIMEDIC™ DefiMonitor XD or by inserting the
PRIMEDIC™ AkuPak battery in the device. The self-test checks all important functions and
signalling devices. Automatically full self-test is initiated in case an internal error is found by
the equipment. Please proceed to follow the device instructions.
6.2 Automatic, periodic self-tests
The device carries out automatic self-tests to ensure that it is always ready for operation. Self-
testing is only possible if the PRIMEDIC™ DefiMonitor XD is either connected to the mains or
if a charged PRIMEDIC™ AkuPak has been inserted.
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 0 V
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 300
V
On the 1st. July and on
the 1st. January each
year
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 1600
V
After identifying an
internal error
Software, operating membrane, ECG calibration,
clock, internal voltage supply and HV part at 1600
V, microphone test