13 Cleaning, maintenance and dispatch
13.1 Cleaning
WARNING
Only clean the device when it is switched off and with the electrodes
unplugged. Before cleaning, remove the power module from the device
or unplug the mains plug!
Do not use any dripping wet cloths to clean it. Never pour liquids over
the device and do not immerse it in any liquids!
Clean the device and all its accessories, such as the wall bracket, with commercially available
household cleaners.
Use a slightly damp, clean cloth.
Use normal disinfectant to disinfect it (e.g. Gigasept FF, Bacillol or Spitacid).
WARNING
Clean the paddles after every use. Soiling as a result of gel residue can
result in danger to the user and patients. Always remove all traces of gel
from the paddles.
13.2 Servicing
Independently of the use of the device, we recommend regular visual inspection / servicing of
the PRIMEDIC™ DefiMonitor XD and the accessories by the user / service technician at least
once a year.
Make sure that the housing, cable, SavePads and all the other accessories are undamaged.
13.2.1 Servicing check list
Check the expiry date
o the SavePads Connect
o the PRIMEDIC™ AkuPaks, and
o if necessary replace the parts.
Check whether
o the status display "OK" is showing.
o you can switch on the device.
o the device automatically carries out the self-test after being switched on.
o the slot for the power supply is clean.
o the device is fully equipped.