On models with SpO
2
measurement, it is possible to show the SpO
2
pulse graph in the log
printout.
Use the integrated cutting edge on the printer cover to tear off the ECG log printout. Tear off
the strip upwards and to the side.
Careful selection of data to be printed out and keeping printing time to a minimum
saves energy and paper and means that the device can be operated for a longer
not on the mains. This applies especially to printing at a speed of 50 mm/s.
If the ECG monitoring is changed during online printing, the printer is stopped immediately, the
header is reprinted and the printout is then continued. Data still stored at the time of the
changeover are discarded, the new printout starts at the time of the changeover.
7.8 Automatic printout after each shock (Autoprint)
The PRIMEDIC™ DefiMonitor XD allows you to record the event automatically each time
defibrillation or cardioversion is performed. Data from 5 seconds before and 5 seconds after
the shock was administered are recorded.
The Autoprint function can be switched on and off in the setup menu. When the device is
delivered, the function is switched off.
If the Autoprint function is activated in the setup menu, it remains active even if the defibrillator
is switched off or the battery replaced.