Operating the pacer
METRAX GmbH
DefiMonitor XD 21324 / GB / K05
77 / 116
Note
Both, the frequency as well as the intensity can be changed during pacing by
pressing the buttons.
During pacing, the ECG alarm is effective provided it was not earlier deactivated in the setup
menu. However, it is not possible to make changes in the setup menu during pacing.
DANGER
To prevent danger to the user as a result of the stimulation pulse, do not
touch the conductive surfaces of the electrodes or the patient during pacing.
After completing the treatment, remove the self-adhesive electrodes from
the patient or ensure that the Pacer / defibrillation cable (SavePadsConnect
cable) is no longer connected to the device.
If the SavePads Connect cable is unplugged from self-adhesive electrodes
or pulled out from the device during pacing, stimulation is immediately
stopped. A warning signal sounds, the cause of the fault is displayed on the
monitor and a voice message is issued:
< Check electrodes>
When the fault has been remedied and the device restarted, the Pacer is
reactivated.
If an energy button is pressed during pacing, pacing is interrupted and a
warning signal sounds to indicate that the process has been interrupted.
The selected energy can thereafter be immediately be loaded.
The patient's ECG must be monitored continually during pacing to ensure
that the stimulation is effective.
The patient must not be left unattended during pacing.
The actual active patient current is monitored. If it falls below 70 % of the set
current, the pulse emission is stopped and a fault message issued.