List of error codes
82 / 116
DefiMonitor XD 21324 / GB / K05
discharge the condenser
64
HV, HV capacitor
overvoltage was detected
Error during the high voltage
load, voltage on condenser too
high
- return the device for servicing
65
HV, HV charger error was
detected
Error during the high voltage
load, impossible to load the high
voltage
- return the device for servicing
66
HV, Error of HV
measurement circuit
Error during measurements in
high voltage section
- return the device for servicing
67
HV, HV charge time is too
big
Error during the high voltage
load, impossible to load the high
voltage
- return the device for servicing
68
HV, Too big current was
detected
Current too high during shock
application
- return the device for servicing
69
HV, Too low current at
shock was detected
Current too low during shock
application, possibly due to
failure in final stage
- return the device for servicing
70
HV, Current sensor error
Error in the current regulation
- return the device for servicing
71
HV, SHKEN net error
Error during the safety test
before the shock application
- return the device for servicing
72
HV, START net error
Error during shock release
- restart the device
- if the error persists: Return the
device for servicing
75
HV, HV hardware
protection error
Error during shock release
- restart the device
- if the error persists: Return the
device for servicing
76
HV, Residual voltage after
disarm error
Residual current to high on the
condensator after the discharge
- restart the device
- if the error persists: Return the
device for servicing
77
HV, Capacitor voltage
does not match energy
level
Wrong condensator voltage with
respect to current/energy
requirements
- return the device for servicing
78
HV, Capacitor voltage is
not constantly increasing
Error during the high voltage
charge, voltage on the
condensator does not increase
in a continuous manner
- return the device for servicing
79
HV, error in calculation of
charge voltage
Error during the high voltage
charge, defective voltage values
for the condensator
- return the device for servicing
80
HV, Shock mode error
Defective switching between
synchronous and asynchronous
shock modality
- restart the device
- if the error persists: Return the
device for servicing
84
PRWIN fail > 100 ms
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
power module
- if the error persists: Return the
device for servicing