Technical Data
90 / 116
DefiMonitor XD 21324 / GB / K05
SpO2 NELLCOR®
Pulse oximetry module
Indication range:
100 %
Calibration range:
100 ... 50 %
Accuracy:
SpO2
Adults 100 ... 70 % +/- 3 digits
New-born babies 95 ... 70 % +/- 3 digits
Contact the manufacturer for further details on test procedures.
Wavelength:
Red: 660 nm, Infrared: 890 nm
Light intensity:
0.5 lumen/cm2
Operating mode:
continuous
Refreshing time:
< 2 sec.
Impedance
measurement
Defibrillation/Pacing:
23 ... 200 Ohm
Monitoring:
500 2500 Ohm
Measurement
frequency:
30 kHz
Measuring precision:
±15 %
Alarms
System:
EKG, SpO2,, defibrillator, power supply, data memory
Physiological:
schockable rhythm
Analysis duration: one
battery
approx. 7 sec to identify schockable rhythm for a battery with 90 %
of the nominal capacity
Monitor
Monitor type:
high-resolution LCD monitor
Monitor size:
115 x 86 mm (diagonal 144 mm, 5.7")
Resolution:
320 x 240 pixels (pixel size 0.36 x 0.36 mm)
Displays:
Heart rate, number of defibrillations, number of detected VF,
reanimation duration, date, time, battery capacity
Defibrillation / pacing
electrodes
SavePads, SavePads
Connect
Pacing: 1h pacing, operation at 120 pulses/min at 140 mA
8h pacing, operation at 60 pulses/min at 70 mA
10 minutes, operation OverDrive pacing at 200
pulses/min at 180 mA