112 PRODIGIT
5.11.1.9. The UUT’s output voltage drop-out lower than the threshold voltage (V-th
Setting), and the OPP trip point is between W_Hi and W_Lo limitation, then
Right 5 digits LCD display will shows "PASS", otherwise shows "FAIL".
5.11.2. Remote control OPP
EX
REMOTE
Set Remote
TCONFIG OPP
Set OCP test
OPP:START 3
Set start load watt 3W
OPP:STEP 1
Set step load watt 1W
OPP:STOP 5
Set stop load watt 5W
VTH 0.6
Set OPP VTH 0.6V
WL 0
Set watt low limit 0W
WH 5
Set watt high limit 5W
NGENABLE ON
Set NG Enable ON
START
Start OPP testing
TESTING?
Ask Testing? 1Testing0Testing End
NG?
Ask PASS/FAIL?0PASS1FAIL
OPP?
Ask OPP watt value
STOP
Stop OPP testing