112 PRODIGIT
5.11.1.9. The UUT’s output voltage drop-out lower than the threshold voltage (V-th
Setting), and the OPP trip point is between W_Hi and W_Lo limitation, then
Right 5 digits LCD display will shows "PASS", otherwise shows "FAIL".
5.11.2. Remote control OPP
EX:
REMOTE
(Set Remote)
TCONFIG OPP
(Set OCP test)
OPP:START 3
(Set start load watt 3W)
OPP:STEP 1
(Set step load watt 1W)
OPP:STOP 5
(Set stop load watt 5W)
VTH 0.6
(Set OPP VTH 0.6V)
WL 0
(Set watt low limit 0W)
WH 5
(Set watt high limit 5W)
NGENABLE ON
(Set NG Enable ON)
START
(Start OPP testing)
TESTING?
(Ask Testing? 1:Testing,0:Testing End)
NG?
(Ask PASS/FAIL?,0:PASS,1:FAIL)
OPP?
(Ask OPP watt value)
STOP
(Stop OPP testing)