84 PRODIGIT
4-6. Remote control command description
4.6.1. PRESET Set and Read the Default of Load
RISE
Syntax [ PRESet] RISE {SP}{NR2}{NL}
[ PRESet] RISE {NL}
Purpose Set and read the RISE SLEW-RATE
Description
1. The definition of RISE SLEW-RATE is load level change or dynamic load can be
programmed of RISE and FALL are completely independent.
2. The value of RISE has to be included the number of the decimal point, otherwise
the command will not be available.
3. The least significant number is the 3th behind the decimal point.
4.3350F will set to the maximum value of the model automatically when the set
RISE is over the specification of Load.
5. The unit is (m)A/uS.
FALL
Syntax [ PRESet] FALL {SP}{NL}
[ PRESet] FALL{NL}
Purpose Set and read the FALL SLEW-RATE
Description
1. The definition of FALL SLEW-RATE is load level change or dynamic load can be
programmed of RISE and FALL are completely independent.
2.3350F will set to the maximum value of the model automatically when the FALL
which has been set is over the specification of Load.
3. The unit is (m)A/uS .
PERI or PERD
Syntax [ PRESet] PERIPERDHIGHLOW{SP}{ NR2}{NL}
[ PRESet] PERIPERDHIGHLOW{NL}
PurposeSet and read the TLOW and Thigh of DYNAMIC when loading
Description
1. A period of loading waveform of DYNAMIC is combined by TLOW and THIGH.
2. The value of TLOW and THIGH have to be included the number of the decimal
Point, otherwise the command will not be available.
3. The least significant number is the 5th behind the decimal point.
4. 3350F will set the value of TLOW or THIGH automatically when the value which
Has been set is over the maximum of the Load.
5. The unit is mS.
LDONv
Syntax [ PRESet] LDONv {SP}{NR2}{NL}
[ PRESet] LDONv{NL}
PurposeSet and Read the voltage of LOAD ON
DescriptionThis command is for setting the Load voltage value of LOAD ON.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Prodigit 3350F Series and is the answer not in the manual?

Prodigit 3350F Series Specifications

General IconGeneral
BrandProdigit
Model3350F Series
CategoryPower Supply
LanguageEnglish

Related product manuals