42 PRODIGIT
Setting the short test timeThe LCD display show SHORT on left 5 digits
LCD displayshows TIMEon middle 5 digits LCD displayright 5 digit LCD
display CONTI, the unit is "ms".
TIME: setting the short test time, The LCD display show SHORT on left 5
digits LCD display, shows TIMEon middle 5 digits LCD display the unit is
“ms” ,and shows CONTI on right 5 digits LCD display, the Setting range is
“CONTI” means continue, 100mS to 10000mS step 100ms by Clockwise rotate the
Setting knob.
The short test will be no time limitation when setting to CONTI until press
“START/STOP” key to stop the short test.
V-Hi : Short test voltage check upper limitation setting, The LCD display shows
SHORTon left 5 digit LCD display, Middle 5 digit LCD displayV-Hi,right 5
Digit LCD display setting value, the unit is "V", The V-Hi setting range from 0.000V to
60.000V step 0.001V by rotating the setting knob.