Communication Protocol CubiScan 25 Command Set
CubiScan 25 Service Manual 67
Save Parameters
This command causes the CubiScan 25 to save all current parameters to the file param125.txt on
the SD card.
Scale Calibration
This function is required when the load cell is replaced, when a new controller is installed, or for
routine calibration. This command causes the CubiScan 25 to reply with a scale calibration code.
Each time this command is sent, the instrument prompts the operator to follow a defined
calibration pattern.
Scale calibration code descriptions:
SA01 Clear the CubiScan 25
SA02 Place test weight on the platform
SA03 Scale calibration complete
Pos Len Description Type Range ASCII
Command Format
1 1 Start of Text Control (STX) 02h
21Save Command Alpha (s) 73h
31End of Text Control(ETX) 03h
4 1 Carriage Return Control (CR) 0Dh
51Line Feed Control(LF) 0Ah
Acknowledge Format
1 1 Start of Text Control (STX) 02h
21Save Command Alpha (s) 73h
3 1 Acknowledge Alpha (A) 41h
41End of Text Control(ETX) 03h
5 1 Carriage Return Control (CR) 0Dh
61Line Feed Control(LF) 0Ah
Negative Acknowledge Format
1 1 Start of Text Control (STX) 02h
21Save Command Alpha (s) 73h
31Neg. Acknowledge Alpha (N) 4Eh
41End of Text Control(ETX) 03h
5 1 Carriage Return Control (CR) 0Dh
61Line Feed Control(LF) 0Ah