Location
TheQuidelTriageMeterProshouldbeplacedonadry,clean,flat,horizontalsurfaceawayfromdirect
sunlight.
Power
TheQuidelTriageMeterProrequires6ͲvoltsDCsuppliedbyeitherACorDC.Themeterwill run
approximately100testsonbatteries.Itisrecommendedtousebatteriesasabackupintheeventofa
poweroutage.
CAUTION:Useonlythepowersupplyprovidedwiththemeter,failuretodosomaycausepermanentdamageto
themeter.
SelfͲTest
TheQuidelTriageMeterProchecksthesystemtobesureit isoperatingproperlybyrunningselfͲtestswhen
poweredonandpriortoeachtest.
PowerOn–includeslaseroperation,internalstandardrange,batterypower,andsoftwareverification.
EachTest–verifiesthelaseroperation,internalstandard
rangeandbatterypower.
CODECHIPModules
MeterdataisupdatedviaadisposableCODECHIPmodule.TheCODECHIPmodulecontainsmicrochip
circuitryembeddedintoaplastichousing.WheninsertedintothemeterCODECHIPmoduleportand
activated,theinformationistransferredintotheQuidelTriageMeterPro’smemory.ACODECHIPmodule
typicallyneedstobe
installedonlyonceandremainsinthemetermemory.ACODECHIPmoduledoesnot
needtoremaininthemeterwhileperf ormingtests.
TherearefivetypesofCODECHIPmoduleswhichcanbedistinguishedbyshapeorcolor.
StandardCODECHIP
module
ReagentTestCODECHIPmodule–includedineach boxoftestdevices;containscalibration,expirationdate
andotherdataaboutthedevicelot.(Colorvariesbytesttype.)
QCSampleCODECHIPmodule–includedineachboxofQCSample;containsexpirationdateandother
data
abouttheQCSamplelotincludingacceptablerange.(White.)
QCDeviceCODECHIPmodule–includedintheblackQCDevicebox(comeswiththeQuidelTriage
MeterPro);containsdataabouthowthemetershouldreaditsmatchingQCDevice.(Darkgray.)
Supervisor AccessCODECHIPmodule–included
witheachQuidelTriageMeterPro;containscodeto
activateSupervisorfunctions.(Lightgray.)
NOTE:TheQuidelTriageMeterProwillautomaticallyturnoffifleftunusedfor2hours,unless
programmedotherwisebythesupervisor.Optionalsettingsallowautomatedturnoffafter30minutes,
1hour,2hours,4hoursorNONE.