lowcontrolsandthebaselinecontrolandtherebyontheimmunoassay,theresultsarenotdisplayedand
erroneousresultsarenotreported.
TheTimingControlZonemonitorsassaycompletion.Ifaninadequateamountofspecimenisaddedtothe
testdeviceorifaspecimenclotsinthedevice,this
controlwillpreventthedisplayoftheresults.
ATraceQCalgorithmhasbeenprogrammedwhichevaluates the size,locationandsignaltonoiseratioof
eachzoneonthedevice.Aberrationsinzonequalityexceedingpresetlimitswillcauserejectionofoneor
moreoftheassayresults.
MeterQualityAssurance
TheSelfͲTestModeisinitiatedeachtimethemeteristurnedON.IntheselfͲtestmodethemeterscansan
InternalCalibrationChip.Eachcalibration chipscanisusedtovalidateandadjust,ifnecessary,themeter
calibration.IntheselfͲtestmode,theavailablememoryforpatientresults
ischecked.Whenthecapacityis
belowapreͲsetlimit,theremainingcapacityisrelatedviaascreenmessage.Thesoftwareprogramis
evaluatedandisterminatedifcorrupted.SoftwaretestresultsandCODECHIPmoduledataareverifiedfor
integritybeforeeachuse.
AQCDeviceisprovided
witheachmetertoallowtheusertofurtherchecktheintegrityofthemeter.The
simulatorcontainssixfluorescentzonesofvaryingintensitythataremeasuredbythemeter.TherearepreͲ
programmedacceptablerangesforthesixzones.Inaddition,thezonesallowthemetertocheckfor
horizontalandverticalalignmentofthezonemeasurementtoverifylaseralignmentanddevicetransport.
TheQCDevicealsoconfirmslaserstability.
ABarCodeisprintedonthebottomofeverytestdevicetoidentifythereagentlotnumberandtoprevent
theuseofexpiredreagents.
SoftwareQualityAssurance
ThesoftwareQAiscontrolledthroughavarietyofsoftwarefeatures,someofwhichcanbeoptional
dependingontheenvironmentinwhichtheassayisperformed.InaPointͲofͲCarelocation,adesignated
supervisorcanselectprogramparameterstoensureadherencetoqualitystandards.Inacentrallaboratory
location,
selectedparameterscanbebypassedifdesired.
ASupervisorCODECHIPmoduleisprovidedwitheachmeter.ThisCODECHIPmoduleallowsaccessto
meterfunctionsnotavailabletotheroutineuser.Theadditionalfunctionsareselectionofparameter
settingsanddownloadingandde letionofresultsfromthemetermemory.
AUserIDcanbeassignedinthesoftwaretolimitaccesstothepatienttestingmodetoonlythose
individualswhohavebeenproperlytrainedandhavedemonstratedtestingproficiency.AUserIDand
expirationdateareprogrammedintothemetermemoryandaresubsequentlyassociatedwithallpatient
resultsgeneratedbytheuser.Thisallowsthesupervisortoreviewusers’testingfrequency,andintheevent
offrequenttestfailures,toidentifypotentialproblemsintraining.
AnadditionalsecurityfeatureoftheUserIDisthepartialmaskingoftheIDondisplaysandprintouts.This
preventsunauthorized
individualsfromdetectingandusinganexistingUserID.Thisfeatureisoverridden
whentheSupervisorCODECHIPmoduleisinstalled.