Page74of86
APatientIDisrequiredforallpatienttestsperformed.All QC sampleandpatientresultsarestoredinthe
meter’smemoryandcanbedownloadedtoalaboratoryinformationsystem(LIS)oranexternaldata
managementsoftwareprogramforfurtherviewingandprintingofqualitycontrolreports.
TheSupervisor
canselecttheQCFreque ncyforperformingexternalliquidcontrolsamples.Whenitistime
toperformexternalQCsamples,theusermusttesttheexternalliquidcontrolsandobtainacceptable
resultsbeforepatientsamplescanbetested.WhenQCsampleshavebeensuccessfullyperformed,theuser
isallowedtoperform
testsonpatientsamples.IfQCsamples arenotwithinanacceptablerange,aQC
Lockoutfunctionpreventstestingofpatientsamples.Ifdesired,thesupervisorinacentrallabsettingcan
bypassthisfunction.Iftestdevicecontrolsareoutofrange,aQCLockoutfunctionisactivatedand
patient
testresultsareblockedfromviewingandprinting.Thispreventsunacceptableresultstobecommunicated
toandusedbythephysician.
Additionally,iftestdevicesareexpired,aQCLockoutfunctionisactivatedandpreventstestingusing
expiredtestdevices.
QualityControlSamples
Itisstillvaluabletoapplythetraditionalapproachtoqualitycontrolbytestingqualitycontrolsamples.
Thesecontrolswillcheckthetotalintegrityofthesystem.Theintervalforanalyzingthesecontrols,
however,canbeextendedduetothemanyother TotalQualityAssurancefeaturesinherentintheQuidel
Triage
tests.
TheQuidelTriagetestshavebeendesignedtomaximizeTotalQualityAssuranceinany testing environment.
ThecombinationoftheQCfeaturesoftheQuidelTriagetestsreducetheimpactofproceduralerrors,
ensurereagentintegrity,andassurethatpatientresultsareaccurateeachandeverytimeatest
is
performed.BasedonthecurrentCLIAguidelinesandotherregulatingbodies,Quidelmakesthefollowing
QCrecommendations:
RuntwolevelsorPOSandNEGexternalliquidcontrolsamplesasa ppropriate witheachnewlotof
reagentsandonceeverythirtydayswithcontinueduseofthesamereagentlot
number
RuntheQCDevicedaily.
NotesRegardingSoftwareControls
1. UserIDAccess
Topreventuntrainedpers ons fromperformingatest,avalidUserID(1Ͳ16charactersinlength)
mustbeenteredbeforeaccessisgiventotheRUN
TESTfunctioniftheUserIDbypassisOFF.Ifa
UserIDBypassisON,thenanyonecananalyzeatest.
Topreventunauthorizedpersonsfromaccessingpatientinformation,avalidUserIDmustbe
enteredbeforeaccessisgiventotheRECALL
PATIENTRESULTSfunctionsiftheUserIDbypassis
OFF.IfaUserIDBypassisON,thenanyonecanrecallpatientdata.RECALL
PATIENTRESULTS
functionsincludere callingpatientresults,printingpatientresults,andrecallingthelasttestrunon
themeterwhenthelasttestisapatientsample,
TheUserIDBypassiscontrolledbytheSET
PARAMETERSfunction,whichrequiresaSupervisor
CODECHIPmoduleforaccess.
AccessisavailabletoanyonewantingtoRECALL
NONͲPATIENTRESULTSfrommemoryorINSTALL
CODECHIP.
2. OutͲofͲRangeFlags