Page72of86
additionalreagents.Eachdevicecontainsafullcompleme ntof reagentsthatarerequiredtoperformthe
immunoassay.
TheQuidelTriagetestshavebeendesignedtoincludeassaycontrolsineachtestdevicethatmonitorthe
assayprocedureandthereagentintegrity.Theassaycontrolspresentauniqueandvaluableextensionto
thetraditionalapproachtoqualitycontrolusingliquidreagents.Iftheproductisstoredunderspecifie d
conditions,
properfunctioningoftheassaycontrolsindicatesthatthetestdeviceisvalidandproper
adherencetoassayprotocolhasbeenfollowed.TheassaycontrolsintheQuidelTriagetestsrequirethe
needtoreͲthinkthefrequencywithwhichexternalqualitycontrolspecimensshouldbetested.
TheQuidelTriage
testshavebeendesignedtoyieldlong Ͳtermreagentstability.Inaddition, thesystem
providesqualitycontrolparametersinthreeseparateanddistinctcategories:
Internaldevicecontrols
Meterelectroniccontrols
Softwarecontrols.
Thecombinationofthesequalitycontrolfeaturesprovides moreinformationaboutthevalidityofeach
assaythancanbeprovidedbythemoretraditionalmethodsofreagentandassayprocedurevalidation.
QualityAssuranceFunctions
1. InternalDeviceQualityAssurance
InternalPositiveControlZone
BaselineControlZone
TimingControlZone
ZoneQualityRequirements(TraceQC)
2. MeterQualityAssurance
SelfͲTestMode
InternalCalibrationChip
ElectronicQCDevice
BarCode
3. SoftwareQualityAssurance
SupervisorCODECHIP
module
UserID#
PatientID#
Resultmemorystorage
QCFrequency
QCLockouts
4. QualityControlSamples
InternalDeviceQualityAssurance
TheInternalControlZonesarebuiltͲin functionalimmunoassays.Thiscontrolverifiesthatthedevice
functionedproperlyandthattheassaywascorrectlyperformed.TheInternalPositiveControlZoneensures
thattheantibodies,thereagentreconstitution,thetimegateandthedeviceflowarefunctioning properly.
Themanufacturersetsacceptablerangesfor
thecontrolzonesforeachlotofreagents.
TheBaselineControlZonemonitorsthepresenceofinterferingsubstancesinthepatientsamplethat may
altertheimmunoassayprocess.Ifaspecimencontainsasubstancethathasamajoreffectonthehighand