www.qutools.com quTAG MC Manual 35
9.4 Low Resolution
Problem
The quTAG MC has very low time resolution.
Cause
After starting the quTAG, the internal electronics are not in a thermal
equilibrium. When the device temperature changes in the first minutes, the
timing jitter degrades due to the different conditions at the startup
calibration.
Solution
Perform the calibration procedure according to the handbook. At the Daisy
start or restart, the calibration procedure is performed automatically.