59
Custom anode/filter calibrations for GE
ange of Application
node/Filt
ub
ilt (um
vl (mm A
addle
Mo/Al GE DMR 22-49 1.0 mm 0.42-0.61 2.2 mm
Rh/Al GE DMR 25-49 1.0 mm 0.42-0.61 2.2 mm
Mo/Mo GE Pristina 22-32 16-44 0.24 -0.51 2.2 mm
Rh/Ag GE Pristina 27-40 12-80 0.27 - 0.78 2.2 mm
Rh/Ag IQST GE Pristina 34 - - No Paddle
Mo/Cu GE Pristina 40-49 250 2.48-3.27 2.2 mm
Rh/Cu GE Pristina 40-49 250 2.42-3.20 2.2 mm
Mo/Mo GE Essential 22-48 16-44 0.24-0.51 2.2 mm
Mo/Rh GE Essential 22-48 6-42 0.20-0.56 2.2 mm
Rh/Rh GE Essential 25-48 12-80 0.27-0.78 2.2 mm
W/Al GE Definium (diag) Same as Diagnostic No Paddle
W/Al GE Optima (diag) Same as Diagnostic No Paddle
Custom anode/filter calibrations for Philips
ange of Application
node/Filt
ub
ilt (um
vl (mm A
addle
W/Al Philips Microdose 20-50 300-3030 0.16-1.96 No Paddle
Custom anode/filter calibrations for Siemens
ange of Application
node/Filt
ub
ilt (
vl (mm A
addle
W/Al
Siemens Straton
Tube CT with W
filter 40-140 0-29 5-10
MDL8155
filter
required
W/Ti
Mammomat
Revelation
41-50
1.0 2.45-3.0 No Paddle
Custom anode/filter calibrations for Hologic
ange of Application
node/Filt
ub
ilt (um
vl (mm A
addle
W/Cu Hologic Dimensions 38-49 174-325 1.54-3.52 No Paddle