94
Appendix E
Accu-Gold measurements and Feature comparison
The following are measurements that can be assigned to any cell when creating custom profiles. These
are grouped by the measurement source. Note that not all measurements are available when using the
Nugget or Touch WiFi mode.
USB Cable/
AGT Pro
Nugget/
Touch WiFi
Definition
Digitizer
Air Pressure yes yes
T
he atmospheric corrections compensate for the effects o
f
variations in the current temperature (T in C) and pressure
(P in kPa) on an unsealed ion chamber and uses the
following equation:
Displayed value = (101.3/P) * ((T + 273.15)/295.15) *
Instrument value
Ave. Pulse Duration yes yes
Average duration of pulses.
Duration yes yes
E
xposure duration from start to stop trigger.
If
pre
-
p
uls
e
detection was activated, duration is the duration of the main
signal without pre-pulse.
Pre-Pulse Detected yes no
T
rue
if
pre
-
p
uls
e
d
etection was active an
d
pre
-
p
uls
e
s
igna
l
was detected.
Pre-Pulse Duration yes no
D
uration from detection o
f
pre
-
p
uls
e
s
tart to main signa
l
start.
Pulse Count yes yes
Measurement pulse count.
Pulse Duty Cycle yes yes
Pulse duty cycle.
Pulse Frequency yes yes
Pulse frequency
Pulse Rep. Period yes no
Measurement pulse repetition period.
Total Duration yes no
E
xposure duration from start to stop trigger, ignore
s
pre
-
pulse detection.
Trigger Level yes yes
Trigger level used for this measurement.
Trigger Sensor yes yes
Sensor that was trigger.
Typical Pulse Count yes no
Measurement pulse count ignoring scout pulse.
Multi-Sensor
Anode yes yes
Tube anode setting used for this measurement.
Average kV yes yes
A
verage kV is derived from the ratios of the outputs o
f
diodes in the multi-sensor, where these averages use the
entire region of interest.
Biased Average kV yes no
T
he average value of kV in the region of interest above a
n
adaptive kV threshold.
Calibration Description yes yes
T
he name of the calibration and correction that was use
d
for making the measurement. e.g. "Diagnostic", "Standard",
"Mo Mo", "Mo Mo GE" etc.
Dose yes yes
Dose Integral
Dose / mAs yes no
Dose / mAs
Dose / Pulse yes yes
Dose / Pulse

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Radcal ACCU-GOLD2 and is the answer not in the manual?

Radcal ACCU-GOLD2 Specifications

General IconGeneral
BrandRadcal
ModelACCU-GOLD2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals