10X6-6M
For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-6M is a dedicated mammography chamber.
The chamber is a world standard for mammography, ready for any mammographic tube track-filter
combination. Extraordinary flat energy response over 10 - 40 keV has been documented in technical
papers and makes corrections unnecessary. It is also recommended for measuring exposure time in Auto
Dose mode. The range of time (pulse width) extends from 10ms to 300s.
C
hamb
er
1
0X
6
-
6
M
M
in Ra
te
2
R
/s
2
0 nGy
/s
M
ax Ra
te
1
0 R
/s
8
8 mGy
/s
M
in Dos
e
1
0
R
1
00 nG
y
M
ax Dos
e
5
9 k
R
5
16 G
y
C
ine Specification
s
N/A
C
alibration Accura
cy
±
4% usin
g
X
-
r
a
y
s
@
3
0
kV
p
a
nd
0.
50
m
m AL HV
L
E
xposure Rate Dependenc
e
±
5%, 0.02 R/min to 600 R/m
in
E
nerg
y
D
ependenc
e
±
5%, 10 keV to 40 ke
V
C
onstructio
n
0
.7 mg/cm2 metalized polyester window; polyacetal exterio
r;
6cm3 active volume; 0.08kg