69
10X6-500 Leakage Measurement Chamber
A single sensor leakage measurement solution for very low-level radiation measurements such as
shielding, leakage, irradiators and environmental that is part of the Accu-Gold family of systems.
Equivalent to the Fluke® 96010A Ion Chamber used in the Fluke® Radiation Leakage Detection System,
this chamber meets 21 CFR 1020.30(k) for leakage measurement requirements.
C
hamb
er
1
0X
6
-
5
00
M
in Ra
te
62
n
R
/s
0
.6
n
G
y
/s
M
ax Ra
te
1
3
0
m
R
/s
1.2
m
G
y
/s
M
in Dos
e
1
nR
0
.01
n
Gy
M
ax Dos
e
45
R
0.4
Gy
C
ine Specification
s
N/A
C
alibration Accura
cy
±
4% using
X
-
r
ays @
1
0
0
kVp an
d
3
.9
6
m
m AL HV
L
E
xposure Rate Dependenc
e
<
5%
u
p to 5 Gy/hr (570 R/h
r)
E
nergy Dependenc
e
±
5%
,
40
-
1
5
0
kV
p
(RQR and RQA Beam Qualitie
s)
C
onstructio
n
T
runcated cylinder, polycarbonate walls and electrode
;
conductive graphite exterior coating; 522 cm
3
active volume;
0.54 kg.