For use with Accu-Gold, the AGMS-D is a Solid-State Dose multi-sensor for Diagnostic Range
Measurements. It is used for single exposure and fluoroscopy. It measures dose, dose-rate, time, kVp with
FFT analysis, Flash HVL, and beam filtration. It features Flash correction of dose for beam quality.
T
yp
e
Anode / Filter Tube kV
Filt
(um)
hvl
(
mm
AI) ±10%
or ±0.2
mm AI
whichever
is greater
R
ad/Fluoro/Denta
l
(AGMS-D)
W/AI General
40
-
1
6
0
±2% (2-24 mm
Al, 40-160 kV)
±2.5% (25-40
mm Al, 50-160
kV)
2-40
mm
1.3-13.5
R
ad/Flouro/Denta
l
(AGMS-D)
D
ose usin
g
AGDM
D
ose Rate usin
g
AGDM
D
ose usin
g
AGDM+
D
ose Rat
e
using AGDM+
Time
1
50 nG
y
>100 Gy,
±5% (17 R
– >11 kR,
±5%)
150 nGy/s – 350
mGy/s, ±5% (1.0
mR/min – 2.4
kR/min, ±5%)
80 nGy - >100
Gy, ±5% (9.1
uR - >11 kR,
±5%)
8
0 nGy/
s
-
1
5
0
mGy/s
±5% (0.55
mR/min - 1.0
kR/min, ±5%)
1 ms – 300 s
± 0.1% + 0.2
ms
Reference condition: 12
° Tungsten anode