Command reference
R&S
®
ZNL/ZNLE
796User Manual 1178.5966.02 ─ 20
Example: 
*RST; CALC:PAR:MEAS 'Trc1'", '"Y-S22'
Select the converted admittance Y <-- S22 as measurement
parameter of the default trace.
CALC:TRAN:COMP S
Convert the converted Y-parameter into an S-parameter.
CALCulate<Chn>:TRANsform:IMPedance:RNORmal <Model>
Selects the theory for the renormalization of port impedances. The selection has an
impact on the conversion formulas for wave quantities and S-parameters.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Model> TWAVes | PWAVes
TWAVes - travelling waves
PWAVes - power waves
*RST:  TWAVes
Example:  See [SENSe<Ch>:]PORT<PhyPt>:ZREFerence
Manual operation:  See "Renormalization According to Theory of" on page 402
CALCulate... (other)
CALCulate:CLIMits:FAIL?............................................................................................... 796
CALCulate<Chn>:DLINe.................................................................................................797
CALCulate<Chn>:DLINe:STATe.......................................................................................797
CALCulate<Chn>:FORMat..............................................................................................797
CALCulate<Chn>:FORMat:WQUType..............................................................................799
CALCulate<Chn>:GDAPerture:SCOunt............................................................................799
CALCulate<Chn>:IAVerage:MODE.................................................................................. 800
CALCulate<Chn>:IAVerage[:STATe].................................................................................800
CALCulate<Chn>:PHOLd............................................................................................... 800
CALCulate<Chn>:SMOothing:APERture.......................................................................... 801
CALCulate<Chn>:SMOothing[:STATe]..............................................................................801
CALCulate:TDVSwr[:STATe]............................................................................................802
CALCulate:CLIMits:FAIL?
Returns a 0 or 1 to indicate whether or not a global, composite limit check on several
traces has failed.
Since V2.20 of the R&S ZNL/ZNLE FW the result is automatically recalculated when-
ever a relevant setting is changed, i.e. a subsequent query will return the updated limit
violation state.
VNA command reference