RT-7600S Auto Hematology Analyzer User’s Manual
2 Installation
2.1 Instrument unpacking
1. Unpack the instrument’s package and remove material for transportation. Please
keep original packing carton and packing material, in case you need to repack the
instrument in the future.
2. Take out the instrument from plastic package.
3. In accordance with packing list, make sure the packing carton content includes:
RT-7600S hematology analyzer
Note: Product packing feature is as per packing list, in case of inconformity,
please contact seller.
2.2 Installation environment
In order to ensure instrument work normally, please choose working place compliant with
following condition to place RT-7600S Auto Hematology Analyzer:
No massive dust or powder;
No strong electromagnetic radiation;
Sufficiently large flat and solid desktop.
Note: Instrument working environment: temperature 15
℃ ~
35
℃
, relative
humidity 10%
~
90%.
2.3 Power requirement
Warning: (1) AC power must be well earthed.
(2) AC power shall be stable, sharing with heavy-duty power
appliance is forbidden, rectified power supply is better
equipped.
(3) If there is smoke, smell or noise in instrument, immediately
shut off power, and contact distributor.
(4) When plugging power line, one must catch the plug itself,
instead of power line.