RT-7600S Auto Hematology Analyzer User’s Manual
25
4.2.1 X-B QC setting
Fig.4-7 X-B QC setting
If X-B analysis is “On”, samples are tested and also taken as X-B QC analysis data.
X-B QC target value and limits are calculated and filled out by user manually. Parameter
target value is the average of statistic calculation result of sample data after test in this region,
samples shall reach a certain amount, its recommended quantity is above 1000, and sample
source shall be verified to be random, data reference values of different regions will be
different. Limits choice is recommended to be 5%-10%.
Save: save QC setting data.
Items: user can select QC of all parameters, or select QC of some parameters.
Delete: delete X-B QC data, including setting data and test data.
Send: use instrument RS-232 serial port to transfer X-B QC setting data to designated
receiver. Before sending, please use RS-232 serial port cable to connect RT-7600S to
equipment to receive data, start up and set up receiver software parameters. For
instrument transfer terminal parameter setting, please refer to Chapter 7 “System
Setting”.
Help: display system help.
Exit: return to main screen.
4.2.2 X-B QC list
The instrument can save at most 2000 X-B QC data, open X-B QC list, as shown below: